APF Electronics Omega temperature controller configuration


Page last updated: April 8, 2010

The temperature controller for the iodine cell is an Omega CNi833-EI. This model has pid control loop, 2 spdt relay contacts, and a network interface. This document is provided to aid in quickly configuring a replaced or new controller. At the begining of the APF project two units were purchased and configured the same so hopefully this document won't ever be needed!

The controller has been mounted into a panel that adds a power switch and both an AC input fuse and an AC output fuse. Under normal circumstances, the power should always be left on to keep the iodine vaporized.

Please refer to the Omega User's Guide for further detail

 

Configuration for APF:

SP1
70
         
SP2
00.0
         
   
         
INPt
Rtd
385.3      
   
         
RdG
dEC
FFF.F      
   
tEMP
°C      
   
FLtR
4      
   
         
ALR1
dSbL
         
ALR2
dSbL
         
LOOP
dSbL
         
OUt1
SELF
         
OUt2
SELF
         
RAMP  
dSbL
         
   
         
Id
CH.Id
0000      
   
FULL
dSbL      
   
Sd.Id
dSbL