9/18/97
DEIMOS Stage Inter-operation
Instrument Rotation |
Grating Clamp |
Slit Insert. |
Filter Wheel |
Grating Slide |
Grating Tilt |
Slit Cass-ette |
STAGE
STAGE |
OK |
OK |
1 |
OK |
OK |
OK |
xxx |
Slit Cassette |
OK |
OK |
OK |
OK |
OK |
xxx |
OK |
Grating Tilt |
OK |
3 |
OK |
OK |
xxx |
OK |
OK |
Grating Slide |
OK |
OK |
OK |
xxx |
OK |
OK |
OK |
Filter Wheel |
OK |
OK |
xxx |
OK |
OK |
OK |
1 |
Slit Insert. |
OK |
xxx |
OK |
OK |
3 |
OK |
OK |
Grating Clamp |
xxx |
OK |
OK |
OK |
OK |
OK |
OK |
Instrument Rotation |
Notes:
1a. Can’t move insert until cassette is in a
discrete position.
b.
Can’t move cassette unless insert is retracted.
(Note 2 removed 9/18/97 per J. Osborne)
3a. Grating clamp can only go on in discrete
positions.
b.
Clamp
must come off before moving slide and must remain off until slide has completed
it’s move.